CPSCMarch 13, 2002consumer

CPSC, Viessmann Manufacturing Company (U.S.) Inc., Announce Recall of SR-RN-2 and SR-RN2R Boiler Control

Official recall recorded by the U.S. Consumer Product Safety Commission.

The hazard

The controls have an electronic relay that may fail in a manner that will allow the burner to operate and by-pass the interlock with the flue damper. The interlock assures proper operation of the damper for removal of combustion gases. If the interlock fails, excess carbon monoxide emissions may occur.

What to do

Consumers who have one of these boilers should contact their local distributor or Ms. Ulli Hiltz, the Viessmann Manufacturing Company contact. Viessmann Manufacturing Company will install a wiring module in suspect boilers which will put a redundant switch in series with the switch that may erroneously close and by-pass the safety interlock with flue damper. Viessmann Manufacturing Company has contacted distributors and wholesalers and made arrangements for them to contact customers and make arrangements to retrofit the Atola-RN Boilers with SR-RN2 or SR-RN2R Controls.

Brands named

viessmann manufacturing company u scpsccpsc viessmann