FDAMarch 19, 2021drug
ChloraPrep With Tint 2% w/v chlorhexidine gluconate (CHG) and 70% v/v isopropyl alcohol (IPA) Non-Sterile Solution - Hi-Lite Orange, 0.10 fl. oz. (3 ml) each, 25 applicators in carton, applicator is sterile if package is intact. CareFusion El Paso, TX 79912, NDC 054365-400-11 Cat. No. 260415
Official recall recorded by the U.S. Food & Drug Administration.
The hazard
CGMP Deviations: Product is being recalled because at labeled storage conditions of 30*C/75% Relative Humidity, there is the potential for growth of Aspergillus penicillioides, a type of fungus, resulting in a breach of the package integrity.
What to do
FDA enforcement status: Terminated
Brands named
carefusion 213carefusion
UPCs
05436540011
Recall history
Other federal recalls tied to the same brand — a factual pattern, straight from the source data.
- FDABD ChloraPrep Clear, (2% w/v chlorhexidine gluconate (CHG) and 70% v/v isopropyl alcohol (IPA) 1 mL Applicator, 60 Applicators, Sterile Solution, Care Fusion 213, LLC, El Paso, TX 79912, subsidiary of Beckton, Dickinson and Co., NDC 54365-400-31.2026-05-28
- FDABD ChloraPrep FREPP Clear,(2% w/v chlorhexidine gluconate (CHG) and 70% v/v isopropyl alcohol (IPA)) 1.5 mL Applicator, 20 Applicators, Sterile Solution, Care Fusion 213, LLC, El Paso, TX 79912, subsidiary of Beckton, Dickinson and Co., NDC 54365-400-30.2026-05-28
- FDABD ChloraPrep FREPP Clear, (2% w/v chlorhexidine gluconate (CHG) and 70% v/v isopropyl alcohol (IPA) 1.5 mL Applicator, 20 Applicators, Sterile Solution, Care Fusion 213, LLC, El Paso, TX 79912, subsidiary of Beckton, Dickinson and Co., NDC 54365-400-30.2026-05-28
- FDABD ChloraPrep Clear (2% w/v chlorhexidine gluconate (CHG) and 70% v/v isopropyl alcohol (IPA)), 1 mL Applicator, 60 Applicators per inner Carton, Sterile Solution, Care Fusion 213, LLC, El Paso, TX 79912, subsidiary of Beckton, Dickinson and Co., NDC 54365-400-31.2026-05-28
- FDABD Pyxis MedStation 4000 REF: 303 Medication cabinet2026-04-30
- FDABD Pyxis MedStation ES REF: 323 Medication cabinet2026-04-30
- FDABD Pyxis Mini Main REF: 349 Medication cabinet2026-04-30
- FDABD Pyxis MedBank Tower REF: 1145-00 Medication cabinet2026-04-30
Own something like this?
AllClear screens your whole list against every CPSC, FDA, USDA and NHTSA recall — and tells you the day a new one names something you own.
Start free — 200 items, no card →