FDAJanuary 15, 2015device

ARCHITECT c4000 Cuvette Segment, List number 02P75-01. The firm name on the label is Abbott Laboratories, Abbott Park, IL. Cuvette segments are a component to the ARCHITECT c4000 system. Cuvette segments are racks that sit in the reaction carousel and hold cuvettes. Each cuvette segment holds 11 ...

Official recall recorded by the U.S. Food & Drug Administration.

The hazard

There is a potential to generate falsely-depressed patient results in the cuvettes adjacent to the detached segment post for any assay run on the c4000 instrument.

What to do

FDA enforcement status: Terminated

Brands named

abbott laboratoriesabbott