FDAJuly 28, 2016device
Carina Mobile Lifts, Models: a) Carina350EM, REF 6060011 b) Carina350EML, REF 60600012 c) Carina350EE, REF 60600009
Official recall recorded by the U.S. Food & Drug Administration.
The hazard
A limited amount of Carina mobile lifts have been delivered with a too small locking pin. If a locking pin smaller than the correct size is mounted on the Carina mobile lift, the locking pin is likely to cause increased wear on the plastic bushings. In case of continued use, these components may cause wear on the metal of the piston rod eye.
What to do
FDA enforcement status: Terminated
Brands named
handicare abhandicare
Recall history
Other federal recalls tied to the same brand — a factual pattern, straight from the source data.
- FDASwiftHook sold as the following: a. Individually as SwiftHook b. Included with Vega505EE mobile lifts Product Usage: The mobile lifts are intended to assist people with functional impairments in scenarios involving physical positioning.2020-08-21
- FDAHandicare C Series Patient Lift Product Usage: A device used to lift and transport a patient from one place to another.2017-11-03
- FDAHandicare P600 Series Patient Lift Product Usage: A device used to lift and transport a patient from one place to another.2017-11-03
- FDAPRISM MEDICAL A-625, MAX. LOAD 625 lbs. 283 Kgs., Patient lift, non-A/C powered2017-09-08
- FDAEVA Floor Lifts, Model Numbers: 400, 450, 6002016-06-15