FDAJune 14, 2023device

IntelliVue Patient Monitor MX400/MX450/MX500/MX550 with software version P.01.01 and ordered with Option CP2 are affected as follows: 1) IntelliVue MX400 866060; 2) IntelliVue MX450 866062; 3) IntelliVue MX500 866064; 4) IntelliVue MX550 866066 Indicated for use by health care professionals whene...

Official recall recorded by the U.S. Food & Drug Administration.

The hazard

Devices with Option CP2 and initially programmed to software version P.01.01 were configured incorrectly at the factory. Therefore, Enhanced ECG Capabilities provided by Option CP2 were not enabled in these devices. Without Option C01 Full Arrhythmia, the device will not provide the yellow alarms for enhanced arrhythmia detection. This is a retrospectively reported recall from 6/14/23.

What to do

FDA enforcement status: Ongoing

Brands named

philips north americaphilipsphilips north

UPCs

008848380387690088483803877600884838038783

Recall history

Other federal recalls tied to the same brand — a factual pattern, straight from the source data.

Own something like this?

AllClear screens your whole list against every CPSC, FDA, USDA and NHTSA recall — and tells you the day a new one names something you own.