FDAAugust 6, 2021device

TECNIS Eyhance IOLs are permanent intraocular implants intended for implantation in the capsular bag. The TECNIS Simplicity Delivery System is designed as a single-use disposable preloaded modular cartridge that functions as both the primary packaging and as a sterile, disposable insertion system...

Official recall recorded by the U.S. Food & Drug Administration.

The hazard

Nonconforming product was distributed in error. Product was nonconforming due endotoxin levels higher than both the Firm's internal endotoxin specification limit and the FDA recommended endotoxin specification limit. Product was released in error based on the results of additional testing performed during the investigation into release testing nonconformance.

What to do

FDA enforcement status: Terminated

Brands named

johnson johnson surgical visionjohnsonjohnson johnson

UPCs

05050474745831900001985625