FDAApril 19, 2023device

Patient Information Center iX, Product Code 866386, and Patient Information Center iX Expand, Product Code 866390, Software Revision 4.0.1

Official recall recorded by the U.S. Food & Drug Administration.

The hazard

During extended operation of the MX40 with the PIC iX, the DHCP (Dynamic Host Configuration Protocol) lease can expire on the MX40. If the DHCP lease expires while the device is offline, upon reconnecting with the PIC iX, a new IP address will be assigned to the MX40. The PIC iX will not recognize the new IP address after reconnecting and therefore the expecting settings sync fails.

What to do

FDA enforcement status: Ongoing

Brands named

philips north americaphilipsphilips north

UPCs

00884838104594

Recall history

Other federal recalls tied to the same brand — a factual pattern, straight from the source data.