FDAAugust 23, 2023device

LFS Flat screen arm system with surgical lamp for Models Aurora series, Aurora 2 series, Aurora 3 series, Aurora 4 series, Aurora Astro series, Stellar XL, and Stellar series, LFS Flat Screen Arm System Model Numbers LFSST2323, LFSLFSST29TV23, LFSST29TV23, LFSST23TV, LFSST23, LFSST2323TV, LFSLFSS...

Official recall recorded by the U.S. Food & Drug Administration.

The hazard

The stopper block on the LFS bracket could be broken or missing. Without this stopper block, the flat screen attached to the LFS flat screen system can rotate beyond the original angle limit and it could cause excessive force to be given to the spring shaft of the LFS bracket. Due to the excessive force given to the spring shaft, the spring shaft could break and shear off.

What to do

FDA enforcement status: Ongoing

Brands named

dkk dai ichi shomeidkkdkk dai

Recall history

Other federal recalls tied to the same brand — a factual pattern, straight from the source data.