FDAMay 30, 2022device

Positron Emission Tomography and Computed Tomography System, Model: uMI 550

Official recall recorded by the U.S. Food & Drug Administration.

The hazard

The wireless VSM module of a mobile PET/CT system, operating in an environment with strong Wi-Fi signals, may experience ECG signal and respiratory signal loss due to Wi-Fi interference. ECG and respiratory signal loss during acquisition can result in the failure of ECG and respiratory-gated reconstruction of the PET scan, which may require rescanning of the patient.

What to do

FDA enforcement status: Ongoing

Brands named

shanghai united imaging healthcareshanghaishanghai united

UPCs

06971576832026

Recall history