FDASeptember 15, 2022device

ULTRASITE US04TL Smallbore T-Port Extension Set, Product Code 474515, 100 units per carton. to provide needle-free access on I.V. and Extension Sets

Official recall recorded by the U.S. Food & Drug Administration.

The hazard

The extension set has a high probability of failure which could lead to possible leakage or detachment at the device's bonded joint during use. The issue may result in delay of therapy, increased risk of bloodstream infections (BSI), and may expose both patients and healthcare workers to hazardous medications. Depending on the nature of the leak and type of infusion, there is also the potential for life-threatening situations to occur including air embolization or hypovolemia.

What to do

FDA enforcement status: Ongoing

Brands named

b braun medical

UPCs

04046964185670006162749131

Recall history