FDASeptember 15, 2022device

BD Vacutainer Tube Sodium Fluoride: 100 mg, Potassium Oxalate: 20 mg, Catalog Number 367001. For collecting, transporting, separating, and processing blood for blood alcohol determination.

Official recall recorded by the U.S. Food & Drug Administration.

The hazard

The affected product contains isobutylene which has recently demonstrated potential interference in determination of methanol by gas chromatography methods. The interference may lead to false positive results in methanol testing.

What to do

FDA enforcement status: Ongoing

Brands named

becton dickinsonbectonbecton dickinson

UPCs

503829036700123038290367001800382903670017

Recall history

Other federal recalls tied to the same brand — a factual pattern, straight from the source data.