FDAJune 7, 2018device

GE Healthcare MRI Systems, Software Version 12.x, sold under the following product names: SIGNA MR/I ECHOSPD W/ EXCITE, 1.5 HISPEED HD 8-CH 400-AP, MR GOLDSEAL 1.5T HD 8CH USED, UPGRADE 1.5 EXC-II TO EXC-III HD, SIGNA MR/I HISPD W/ EXCITE, 1.5 ECHOSPEED HD 16-CH 800-AP, 1.5 HISPEED HD 4-CH 200-AP...

Official recall recorded by the U.S. Food & Drug Administration.

The hazard

The latest on-site software version may not have been reinstalled at some sites after service activities were performed resulting in images possibly flipped left/right and/or there can be patient data mismatch.

What to do

FDA enforcement status: Terminated

Brands named

ge medical systems

UPCs

000200504115002005120135002004102804002004091301002005101815002002060511002005011028000200508237002006012710002005051113000009901112002005030106