FDAAugust 27, 2015device

MRIdian¿ ViewRay¿ Radiation Therapy System, ViewRay¿ Is indicated for stereotactic radiosurgery and precision radiotherapy for lesions, tumors, and conditions anywhere in the body when radiation treatment is indicated.

Official recall recorded by the U.S. Food & Drug Administration.

The hazard

ViewRay discovered that in the event that an encoder breaks or fails on the Patient Handling System (PHS, or couch), when attempting to restart the system, the couch could move unexpectedly.

What to do

FDA enforcement status: Terminated