FDASeptember 26, 2023device

THERASPHERE ADMIN SET, Article No. 7M2650. Canada only. Sterile, single use administration set used to deliver Yttrium-90 glass microspheres for internal radiation therapy.

Official recall recorded by the U.S. Food & Drug Administration.

The hazard

Devices may leak as a result of a manufacturing issue. Leakage may result in a delay of therapy, increase risk of bloodstream infection (BSI), and may expose both patients and healthcare workers to hazardous medication. There is also the potential for life-threatening situations to occur requiring additional medical intervention.

What to do

FDA enforcement status: Ongoing

Brands named

b braun medical

UPCs

5060116920635

Recall history

Other federal recalls tied to the same brand — a factual pattern, straight from the source data.

Own something like this?

AllClear screens your whole list against every CPSC, FDA, USDA and NHTSA recall — and tells you the day a new one names something you own.

Start free — 200 items, no card →
THERASPHERE ADMIN SET, Article No. 7M2650. Canada only. Sterile, single use administration set used to deliver Yttrium-90 glass microspheres for internal radiation therapy. — Recall Details · AllClear