FDASeptember 14, 2016device

Agee-WristJack¿ Fracture Reduction System; Item number: CFD-147. The WristJack Fracture Reduction System is a fracture reduction and external fixation system for treatment of distal radius fractures. This is a single use system used for a trauma related surgery The WristJack Fracture Reduction Sy...

Official recall recorded by the U.S. Food & Drug Administration.

The hazard

A number of Agee WristJack¿ Fracture Reduction Systems have been manufactured with a molded beam slightly larger than optimal. This may result in the surgeon experiencing difficulty fine-tuning the length (red) and radial/ulnar alignment (blue) adjustments on the device.

What to do

FDA enforcement status: Terminated

Brands named

hand biomechanics labhandhand biomechanics

Recall history

Other federal recalls tied to the same brand — a factual pattern, straight from the source data.

Own something like this?

AllClear screens your whole list against every CPSC, FDA, USDA and NHTSA recall — and tells you the day a new one names something you own.

Start free — 200 items, no card →
Agee-WristJack¿ Fracture Reduction System; Item number: CFD-147. The WristJack Fracture Reduction System is a fracture reduction and external fixation system for treatment of distal radius fractures. This is a single use system used for a trauma related surgery The WristJack Fracture Reduction Sy... — Recall Details · AllClear