FDASeptember 23, 2015device

TELETOM EQUIPMENT MANAGEMENT SYSTEM 4 and 6 series TELETOM is designed for use in operating rooms, medical treatment rooms, intensive care units, or any area of a medical facility that requires a ceiling mounted device to supply power, medical gases, and data-communication services. TELETOM is de...

Official recall recorded by the U.S. Food & Drug Administration.

The hazard

During the use of the TELETOM Series 4 & 6 there is potential for the controls and brakes not to work properly due to fluid intrusion within the control switches and/or the control board assembly.

What to do

FDA enforcement status: Terminated