FDASeptember 16, 2015device

CC cartridges for IRMA TRUPOINT Blood Gas Analyzer, PN 039903. The CC Cartridge are single use, disposable, in vitro diagnostics intended for point of care (POC), professional, use with the IRMA TRUpoint¿ Blood Analysis System.

Official recall recorded by the U.S. Food & Drug Administration.

The hazard

May report values outside the published performance specifications for pH. After a certain point in shelf life, the pH sensor may give results that are falsely elevated. The discrepancy is greater as shelf life progresses and can be seen on the aqueous controls before the blood.

What to do

FDA enforcement status: Terminated