FDASeptember 21, 2022device

WEB Detachment Controller, REF: WDC-2, WDC, MV-WDC2. Component of WEB Aneurysm Embolization System.

Official recall recorded by the U.S. Food & Drug Administration.

The hazard

Detachment controller, of an aneurysm embolization system, has an out of specification tunnel opening dimension, which may prevent the implant device pusher from being fully inserted into the controller, which may result in the inability of the controller to properly detach the implant device, which could cause delays and excessive manipulation that may cause the implant to move and damage vessels

What to do

FDA enforcement status: Ongoing

Brands named

microvention

UPCs

008424291031280084242911368404987892122231