FDASeptember 15, 2023device

Senhance Surgical System, which includes: Manipulator Arm, ISU Configuration ( Catalog Number X9007708), Intelligent Surgical Unit (Catalog Number X9007696), and Cockpit, ISU Configuration (Catalog Number X9007707), Software (SW) version 2.7.4; robotic surgical system

Official recall recorded by the U.S. Food & Drug Administration.

The hazard

Asensus Surgical has identified a malfunction that has occurred and may recur on the Senhance Surgical System. This issue presented itself as uncontrolled arm motion of the Laparoscope Instrument Actuator (LIA) where the LIA rotated continuously in one direction after the Surgeon removed the engagement of teleoperation on the Senhance System. There was no other uncontrolled arm motion observed.

What to do

FDA enforcement status: Terminated

Brands named

asensus surgicalasensus

UPCs

00815440022075

Recall history

Other federal recalls tied to the same brand — a factual pattern, straight from the source data.