BRAINLAB EXACTRAC VERO, Model/Catalog Numbers: 46228 EXACTRAC VERO 3.5 46238 EXACTRAC VERO 3.5 46216 EXACTRAC VERO SW UPDATE 3.5.2 TO 3.5.3 46218 EXACTRAC VERO SW UPDATE 3.5.3 TO 3.5.4 The ExacTrac Vero system is intended to be used in conjunction with the MHI-TM2000 Radiation Therapy Linear Acce...
Official recall recorded by the U.S. Food & Drug Administration.
The hazard
Brainlab has internally detected that under specific conditions ExacTrac Vero may not correctly account for this ring angle correction during the calculation of the corresponding couch shift. This may result in the treatment couch not moving to the exact planned position resulting in a deviation between the planned and the treatment target position. This deviation may not be clearly visible to the user if no verification image of the patient position is acquired after this couch shift has been performed.
What to do
FDA enforcement status: Terminated
Brands named
UPCs
Recall history
Other federal recalls tied to the same brand — a factual pattern, straight from the source data.
- FDAOrigin Data Management software versions 3.1.0, 3.1.1, 3.1.2, 3.2.0, 3.2.12025-02-10
- FDACirq Arm System 2.02023-03-02
- FDABrainlab ExacTrac Dynamic software, Medical Charged-Particle Radiation Therapy System, model 20910-01E and 20910-01F.2023-02-14
- FDAExacTrac Dynamic software model numbers: 20910-01, ETD Positioning and Moni. 1.0.0 INST; 20910-01A, ETD Positioning and Moni. 1.0.1 INST; 20910-01B, ETD Positioning and Moni. 1.0.2 INST; and 20910-01C ETD Positioning and Moni. 1.0.3 INST.2022-02-25
- FDAExacTrac Dynamic software, Model 20910-01B ETD Positioning and Monitoring Installer 1.0.2 - Product Usage: intended to position patients at an accurately defined point within the treatment beam of a medical accelerator for stereotactic radiosurgery or radiotherapy procedures, to monitor the patie...2021-05-14
- FDAExacTrac Dynamic (Medical Charged-Particle Radiation Therapy System).2021-03-01
- FDAUltrasound Navigation Software 1.0 and Ultrasound Integration Software for Cranial/ENT Navigation 3.0.x and 3.1.x - Product Usage: intended to be an intra-operative image guided localization system to enable minimally invasive surgery.2020-08-31
- FDACirq Arm System - Model SBS 1.3.2 (Art.No. 56000-01B) and SBS 1.4 (Art.No. 56000-01C) of the CIRQ Arm System (56000 and 56000A)2020-08-20
Own something like this?