FDASeptember 1, 2023device

Breast Plastics Pack, REF BBPL20D

Official recall recorded by the U.S. Food & Drug Administration.

The hazard

During an internal investigation, ACS identified that the components, I.V. Cath (part number 4252535-02) and Adhesive Dermabond (part number DHVM12) were inadvertently subjected to the ACS sterilization process and exposed to EO and higher temperatures than approved by the component manufacturer. As a result, the components may (1) have a loss/lack of functionality, (2) loss of drug efficacy, if applicable, and (3) have higher than specified EO residuals.

What to do

FDA enforcement status: Ongoing

Brands named

american contract systemsamericanamerican contract

UPCs

00191072166241