Medical convenience kits Item Number/Description ANCV78BC HEART PACK - 205947 HGCV05K CV SET UP PACK MHPP97AR PACEMAKER PACK- 206010 UDBH37AW BASIC HEART PACK - 206044 UHPP82AH PERIPHERAL PACK - 209533 UICD62AN CARDIOVASCUALR SUPPLY (PS 907285) UICT16F CARDIO THORACIC DRAPE PACK (PS 907284) UIVA7...

Official recall recorded by the U.S. Food & Drug Administration.

The hazard

Identified products were re-gassed (sterilized more than once) after a nonconformance occurred during the initial Ethylene Oxide (EO) gas injection process. These products have not been validated for exposure to multiple sterilization cycles. Therefore, product quality, safety, and effectiveness cannot be confirmed for product that was re-gassed.

What to do

FDA enforcement status: Ongoing

Brands named

american contract systemsamericanamerican contract

UPCs

191072214362191072235169191072212474191072237446191072235299191072216434191072200495191072150431191072229052191072235602

Recall history

Other federal recalls tied to the same brand — a factual pattern, straight from the source data.

Own something like this?

AllClear screens your whole list against every CPSC, FDA, USDA and NHTSA recall — and tells you the day a new one names something you own.

Start free — 200 items, no card →
Medical convenience kits Item Number/Description ANCV78BC HEART PACK - 205947 HGCV05K CV SET UP PACK MHPP97AR PACEMAKER PACK- 206010 UDBH37AW BASIC HEART PACK - 206044 UHPP82AH PERIPHERAL PACK - 209533 UICD62AN CARDIOVASCUALR SUPPLY (PS 907285) UICT16F CARDIO THORACIC DRAPE PACK (PS 907284) UIVA7... — Recall Details · AllClear