FDASeptember 4, 2025device

Medical convenience kits Item Number/Description BFCT04X CATARACT PACK - 205986 CMEY17AI EYE PACK - 213187 CMEY17AI EYE PACK - 213187 CMEY17AI EYE PACK - 213187 PERP25AQ PEI RETINAL PACK-206025 RFEY50AC EYE PACK-206026 UIVT47AS VITRECTOMY SUPPLY

Official recall recorded by the U.S. Food & Drug Administration.

The hazard

Identified products were re-gassed (sterilized more than once) after a nonconformance occurred during the initial Ethylene Oxide (EO) gas injection process. These products have not been validated for exposure to multiple sterilization cycles. Therefore, product quality, safety, and effectiveness cannot be confirmed for product that was re-gassed.

What to do

FDA enforcement status: Ongoing

Brands named

american contract systemsamericanamerican contract

UPCs

191072225955191072211514191072199720191072202918191072224569