FDAJuly 6, 2020device

9" (23 cm) Appx 0.63 ml, Smallbore Bifuse Ext Set w/2 MicroClave Clear, NanoClave, 3 Clamps, 0.2 Micron Filter, Rotating Luer; 1 unit per pouch, 50 multivac pouches per case UDI:(01)10887709052953(17)241201(30)50(10)4536611;(01)10887709052953(17)250101(30)50(10)4557492 The Intravascular Administr...

Official recall recorded by the U.S. Food & Drug Administration.

The hazard

Identification of a potential manufacturing defect on the internal surface of the NanoClave within specific lots of NanoClave sets, which may inhibit a proper seal with the NanoClave spike.

What to do

FDA enforcement status: Terminated

Brands named

icu medicalicu

UPCs

10887709052953

Recall history

Other federal recalls tied to the same brand — a factual pattern, straight from the source data.

Own something like this?

AllClear screens your whole list against every CPSC, FDA, USDA and NHTSA recall — and tells you the day a new one names something you own.