SIEMENS N/T Protein Control SL for use on the BN II, BN ProSpec Systems: N/T Protein Control SL/L: REF/Product Code OQIN19, SMN 10446079 (Low); N/T Protein Control SL/M, REF/Product Code OQIO19, SMN 10446085 (Mid); N/T Protein Control SL/H, REF/Product Code OQIP19, SMN 10446089 (High);

Official recall recorded by the U.S. Food & Drug Administration.

The hazard

Siemens Healthcare Diagnostics has confirmed a negative bias of approximately -20% for results obtained with N Antiserum to Human Ceruloplasmin when calibrating Ceruloplasmin with various lots of N Protein Standard SL. N/T Protein Controls SL for Ceruloplasmin are similarly biased. --- The affected standards and controls for the affected lots were incorrectly value assigned, which led to a lower result compared to correctly assigned standards. Note: N Antiserum to Human Ceruloplasmin (REF OUIE) works as intended. --- Siemens investigation indicated that N Protein Standard SL and corresponding N/T Protein Control SL (Levels Low, Mid and High) are inaccurately assigned for Ceruloplasmin.

What to do

FDA enforcement status: Terminated

Brands named

csl behringcsl

Recall history

Other federal recalls tied to the same brand — a factual pattern, straight from the source data.

Own something like this?

AllClear screens your whole list against every CPSC, FDA, USDA and NHTSA recall — and tells you the day a new one names something you own.

Start free — 200 items, no card →
SIEMENS N/T Protein Control SL for use on the BN II, BN ProSpec Systems: N/T Protein Control SL/L: REF/Product Code OQIN19, SMN 10446079 (Low); N/T Protein Control SL/M, REF/Product Code OQIO19, SMN 10446085 (Mid); N/T Protein Control SL/H, REF/Product Code OQIP19, SMN 10446089 (High); — Recall Details · AllClear