FDANovember 2, 2017device
pmi, pro med instruments, DORO(R) Sterile Disposable Skull Pins, 3 pieces (3 pieces per blister, 12 blisters per box), STERILE R, Rx ONLY The DORO(R) Skull Pins are inserted into the receptacles of the skull clamp and fixate the head during surgery.
Official recall recorded by the U.S. Food & Drug Administration.
The hazard
The firm received reports of two breakages of the pin tips which occurred during two identical surgery procedures. The breakage occurred at the very distal end of the conical part of the skull pin tip.
What to do
FDA enforcement status: Terminated
Brands named
pro med instrumentspropro med
Recall history
Other federal recalls tied to the same brand — a factual pattern, straight from the source data.
- FDAAdvance Aluminum Chloride Gel Clear Hemostatic Gel Kit, Catalog number 504600-282026-05-27
- FDAeTRAX Needle Sensor - 12G (for Aurora Trackers), Part Number 667-1562026-03-02
- FDAeTRAX Needle System Starter Kit 18G (for Aurora Trackers), Part Number 667-1522026-03-02
- FDAeTRAX Needle Sensor - 14G(for Aurora Trackers), Part Number 667-1572026-03-02
- FDAeTRAX Needle System Starter Kit 16G(for Aurora Trackers), Part Number 667-1512026-03-02
- FDAeTRAX Needle Sensor - 16G(for Aurora Trackers), Part Number 667-1582026-03-02
- FDAeTRAX Needle System Starter Kit 14G (for Aurora Trackers), Part Number 667-1502026-03-02
- FDAeTRAX Needle Sensor - 18G (for Aurora Trackers), Part Number 667-1592026-03-02