FDASeptember 20, 2022device

Insulin Pump/Model: MiniMed 630G/ MMT-1715, MMT-1755, MMT-1754; MiniMed 670G/ MMT-1780, MMT-1781, MMT-1782, MMT-1760, MMT-1761, MMT-1762, MMT-1740, MMT-1741, MMT-1742

Official recall recorded by the U.S. Food & Drug Administration.

The hazard

Medtronic identified a cybersecurity vulnerability in the MiniMed 600 series Insulin Pump Systems associated with the communication protocol that could allow unauthorized access to the pump system. This unauthorized access could be used to deliver too much or too little insulin through delivery of an unintended insulin bolus or because insulin delivery is slowed or stopped, which could lead to hypoglycemia or hyperglycemia. The Remote Bolus feature on insulin pumps should be turned off to prevent the unauthorized access.

What to do

FDA enforcement status: Ongoing

Brands named

medtronic minimedmedtronic

UPCs

6430000000007630000725206431698738340064316965684064316975272664316987384164316987864876300017963200643169752726643169976344643169949706643169939202