ICU Mednet(TM) Medication Management Suite 6.1 and 6.21, Product Code/List Numbers: (a) 16037-64-01; MedNet 6.1 b) 16037-64-02; MedNet 6.1 c) 16037-64-03; MedNet 6.1 d) 16037-64-04; MedNet 6.21 The ICU Medical MedNet Medication Management Suite (MMS) is intended to facilitate networked communicat...

Official recall recorded by the U.S. Food & Drug Administration.

The hazard

Issue 1: The MedNet Meds 6.1 and 6.21 programs, under certain conditions, can change the piggyback medication entry set settings for existing defined medication entries. This includes (a) whether a medication may be piggybacked and (b) whether an infusion may be interrupted by a piggyback. This only occurs when: 1. Medication entries are built with the same displayed name and diluent amount AND; 2. The piggyback or interruptible settings differ from one CCA to another AND; 3. The drug library is exported then imported into the MedNet Meds program. The changed drug library may then be finalized and downloaded to infusers for use in clinical settings. The changes are made without any notice provided to either the MedNet Meds user (pharmacist) or to the clinician operating the infuser that the piggyback or interruptible settings have been altered. Issue 2: The MedNet Meds 6.1 program, under certain conditions different from those described above in Issue 1, can change the piggyback

What to do

FDA enforcement status: Terminated

Brands named

icu medicalicu

Recall history

Other federal recalls tied to the same brand — a factual pattern, straight from the source data.

Own something like this?

AllClear screens your whole list against every CPSC, FDA, USDA and NHTSA recall — and tells you the day a new one names something you own.

Start free — 200 items, no card →