Accu-Chek Connect diabetes management software app, Catalog number 07562462001 / GTIN number 00365702700000, Catalog number 07250452001 / GTIN number 00365702700017 The Accu-Chek Connect Diabetes Management App is indicated as an aid in the treatment of diabetes. The software provides for electro...
Official recall recorded by the U.S. Food & Drug Administration.
The hazard
Certain iOS and Android App versions contain a program error (bug) in the Bolus Advisor feature. After pairing a meter with the app for the first time, a customer may encounter the rare condition in which the countdown timer is not displayed and correction bolus advice is not available for the most recent, valid glucose reading. This same blood glucose value may become available for bolus advice calculation at a later time (countdown timer is displayed).
What to do
FDA enforcement status: Terminated
Brands named
UPCs
Recall history
Other federal recalls tied to the same brand — a factual pattern, straight from the source data.
- FDAPovi-One, 10% Povidone-Iodine Oral Antiseptic, Packaged by Elevate Oral Care, LLC, 346 Pike Road, Suite 6, West Palm Beach, FL 33411, NDC 57511-0611-1.2026-06-04
- FDABicarby" Dialysate; Model Numbers: RFP-400-G, RFP-401-G, RFP-402-G, RFP-404-G, RFP-407-G, and RFP-456-G;2026-05-12
- FDABicarby Dialysate; Model number: RFP-400-G;2026-05-07
- CPSCAdult Portable Bed Rails Recalled Due to Risk of Serious Injury or Death from Entrapment and Asphyxiation; Violate Mandatory Standard for Adult Portable Bed Rails; Imported by Hepo Care Medical Equipment Online2026-04-30
- FDAAvanta Multi-Patient Administration Tube Sets (AVA 500 MPAT (Part 2 of 2)) used with the Medrad¿ Avanta Injection System; Catalog number: AVA 500 MPAT (Part 2 of 2); Affected material numbers: 87629007, 60729458, 86566648;2026-04-24
- FDADELFLEX, Dextrose Peritoneal Dialysis Solution with attached stay-safe Exchange Set for Intraperitoneal Dialysis Only, 2.5% DEX. LM/LC, 2L 5PK, Part Number 054-20222, Fresenius Medical Care North America, 920 Winter Street, Waltham, MA 02451.2026-04-06
- FDADELFLEX, Dextrose Peritoneal Dialysis Solution with attached stay-safe Exchange Set for Intraperitoneal Dialysis Only, 1.5% DEX. LM/LC, 2L 5PK, Part Number 054-20221, Fresenius Medical Care North America, 920 Winter Street, Waltham, MA 02451.2026-04-06
- FDATandem pumps are battery-operated infusion pumps capable of both basal and bolus delivery of insulin. The pumps utilize a motor-driven mechanism to deliver insulin from within a disposable cartridge, through an infusion set, into a patient s subcutaneous tissue. As with current insulin infusion p...2026-01-30