FDASeptember 29, 2022device

Canon Aplio ultrasound system, Model: Aplio i900 TUS-AI900, Aplio i800 TUS-AI800, Aplio i700 TUS-AI700, Aplio a550 CUS-AA550, Aplio a450 CUS-AA450

Official recall recorded by the U.S. Food & Drug Administration.

The hazard

After completion, ultrasound system's Stress Echo may not terminate if user does not manually terminate the protocol. During next patient exam, an abnormal confirmation message will be displayed "Same view was saved in this phase. Do you use this image with StressEcho Review?" If "no", acquired images not saved, and if "yes" images assigned to the previous patient; could cause treatment delay.

What to do

FDA enforcement status: Ongoing

Brands named

canon medical systemcanoncanon medical

UPCs

0498767010252304987670102516049876701025090498767010335304987670103346

Recall history