FDAMay 30, 2018device

Nuclear Magnetic Resonance Imaging Systems that are indicated for use as a diagnostic imaging modality that produces cross-sectional transaxial, coronal, sagittal, and oblique images that display anatomic structures of the head or body.

Official recall recorded by the U.S. Food & Drug Administration.

The hazard

When using respiratory gating with Steady Mode ON in Spin Echo sequence scanning, the SAR (Specific Absorption Rate) value displayed on the console may be lower than the actual SAR value. The actual applied SAR values may exceed the FIRST LEVEL CONTROLLED OPERATING MODE.

What to do

FDA enforcement status: Ongoing

Brands named

canon medical systemcanoncanon medical