FDAOctober 13, 2022device

VITROS Chemistry Products Calibrator Kit 9 -(GEN) 22 of VITROS PHYT Slides For in vitro diagnostic use only. Product Code: 8568040 . VITROS Chemistry Products Calibrator Kit 9 is used to calibrate VITROS 250/350/5,1 FS/4600/XT 3400 Chemistry Systems and the VITROS 5600/XT 7600 Integrated Systems ...

Official recall recorded by the U.S. Food & Drug Administration.

The hazard

Calibration disk Data Release Version (DRV) 6194 to 6200 contain inappropriate calibration mathematics for VITROS Calibration Kit 9 for calibrating VITROS PHYT Slides, for all lots of slide generation 22 (GEN 22). The erroneous mathematics on the calibration disk prevents the VITROS 250/350 Chemistry System from successfully calibrating, leading to delayed results.

What to do

FDA enforcement status: Ongoing

Brands named

ortho clinical diagnosticsorthoortho clinical