FDAOctober 20, 2022device

NEO-fit Neonatal Endotracheal Tube Grip, Part Number 42-2540

Official recall recorded by the U.S. Food & Drug Administration.

The hazard

Firm has received 11 complaints involving a loose or detached metal clips, including 3 cases where the metal clip was ingested by the patient, and 3 cases where the metal clip was removed from the patient's mouth. In one case, the patient suffered a laceration to the throat.

What to do

FDA enforcement status: Ongoing