FDAOctober 5, 2021device

Sensis/ Sensis Vibe systems with software VD1X as follows: (1)Sensis Model 10764561 (2) Sensis Vibe Hemo Model 11007641 (3)Sensis Vibe Combo Model 11007642 Intended Use: A diagnostic and administrative tool supporting hemodynamic catheterizations and/or electrophysiology studies

Official recall recorded by the U.S. Food & Drug Administration.

The hazard

System operator manual states that the system should be rebooted once, every 7 days. In some cases, users are not following these instructions and the system is running for more than 7 days. Under rare circumstances, this may result in a partial freeze of the user interface which would no longer update the vital signs

What to do

FDA enforcement status: Ongoing

Brands named

siemens medical solutionssiemenssiemens medical