FDASeptember 18, 2024device

a. Hillrom Centrella Smart+ Medical-Surgical Hospital Bed, Product codes: P7900B000001, P7900B100010, P7900B100024, P7900B100208, P7900B100273, P7900B100304, P7900B000005, P7900B100011, P7900B100025, P7900B100212, P7900B100277, P7900B100307, P7900B000010, P7900B100017, P7900B100026, P7900B100222,...

Official recall recorded by the U.S. Food & Drug Administration.

The hazard

The beds have a potential for an interface issue with the Rauland Responder 5 Nurse Call System where the bed exit system will alert at the bed, but may fail to send a remote alert through the nurse call system.

What to do

FDA enforcement status: Ongoing

Brands named

baxter healthcarebaxter

UPCs

0088776198516200887761999961