FDAOctober 13, 2022device

BD Phoenix PMIC/ID-107: The Phoenix PMIC/ID panel is used for the rapid identification and susceptibility of most aerobic and facultative anaerobic gram-positive bacteria of human origin with selected antimicrobial agents when used with the BD Phoenix Automated Microbiology System. Catalog (Ref) ...

Official recall recorded by the U.S. Food & Drug Administration.

The hazard

Misformulated minocycline-can cause erroneously increased Minimal Inhibitory Concentration (MIC) values when performing antibiotic susceptibility testing (AST) against minocycline and result in the reporting of false resistance for clinical samples if the initial Quality Control (QC) testing of the panel batch passes, delaying treatment of the patient with the most appropriate antibiotic

What to do

FDA enforcement status: Ongoing