FDAAugust 5, 2025device

t:slim X2 insulin pump with Interoperable Technology, Control-IQ+ technology, REF: 1014070, 1014071, 1014072, 1014073, and User Guide when using Dexcom G7 sensor; Tandem Mobi insulin pump with Interoperable Technology, Control-IQ+ technology, REF: 1014801, 1014811, 1014813, and User Guide when us...

Official recall recorded by the U.S. Food & Drug Administration.

The hazard

A software defect in versions 7.9.0.1 and 7.10.1 of the pump software for Tandem t:slim X2 and version 7.9.0.1 of the Tandem Mobi pump with Control-IQ+ technology that are paired with a G7 sensor may result in an unexpected automatic insulin correction bolus (autobolus ) which could result in hypoglycemia.

What to do

FDA enforcement status: Ongoing

Brands named

tandem diabetes caretandemtandem diabetes

UPCs

00389152407012003891524071110038915240721000389152407319003891524801140038915248111100389152481319