VariSoft infusion set, Model Number 1002830
Official recall recorded by the U.S. Food & Drug Administration.
The hazard
The Varisoft infusion set is used in conjunction with tandem insulin pumps. The defect reported was damage to the connector piece during manufacturing. This damage results in the connector becoming detached more easily from the insulin set than expected, requiring less force to disconnect than intended. Disconnection, when not detected, for example during sleep, leads to missed basal dosing. which in turn can result in elevated blood glucose and ketone level (nocturnal hyperglycemia). Diabetic ketoacidosis.
What to do
FDA enforcement status: Ongoing
Brands named
UPCs
Recall history
Other federal recalls tied to the same brand — a factual pattern, straight from the source data.
- FDAVariSoft Infusion Set, Single Use.2024-07-23
- FDANeria Soft Infusion Set, Product Code 507302, intravascular administration set2024-03-05
- FDAVariSoft infusion set, Model Number 10028282023-10-11
- FDAVariSoft infusion set, Model Number 10028272023-10-11
- FDAAutosoft 90, model numbers 1000282, 1002817, 1002818, 1002819, 1002820, 1002821, 1002822, 1002823, and 1002824. Infusion sets used with Tandem Insulin Subcutaneous Infusion Pump.2022-07-21
- FDASure T, contact, contact detach, neria, neria detach, neria multi and thalaset Used in combination with infusion pumps for continuous subcutaneous infusion of medication.2015-05-21
- FDAInfusion Set, Comfort, 23" Hosp 10; Comfort 13mm, 23", STD 5/5; Comfort 13mm, 23" Hosp 10; Infusion Set, Comfort, 43", Hosp 10; Infusion Set, Comfort, 31", Hosp 10; Comfort 13mm, 31". Hosp 10; Infusion Set, Comfort, 23", STD 5/6; Comfort 13mm, 43" Hosp 10; Infusion Set, Comfort 43", STD 5/5; Infu...2014-11-07
- FDAContact Detach, 8mm, 23", G29; Contact Detach 8 mm, 43", G29; Contact Detach, 6mm, 23" G29; Infusion Set, Contact Detach, 6mm x 23"; Infusion Set, Contact Detach, 6mm x 32"; Infusion Set, Contact Detach, 8mm x 23"; Infusion Set, Contact Detach, 8mm x 32" The products are used in combination with ...2014-11-07