FDASeptember 6, 2018device

Altrix¿ Precision Temperature Management System Model: 8001Product Usage: The Altrix system is intended for circulating temperature controlled warm or cold water via patient contact thermal transfer devices for the application of regulating human body temperature in situations where a physician o...

Official recall recorded by the U.S. Food & Drug Administration.

The hazard

Users may experience alarm fatigue due to frequent alarming relating to patient temperature deviation beyond 0.5¿C while in Automatic mode. 2. Users may experience a Remove from Use code 9 (RFU 9) fault condition relating to inadvertent flow alarms.3. Users may experience code RFU 27 fault condition relating to mechanical interference between the device fan and filter

What to do

FDA enforcement status: Terminated

Brands named

stryker medical divisionstrykerstryker medical