FDAOctober 1, 2019device

Accutorr 3 Vital Signs Monitor with NIBP valve assembly board, p/n 115-017679-00, containing software version 8.1.5 - Product Usage: The monitor is intended for spot-check monitoring physiologic parameters, including Pulse Oximetry (SpO2), Pulse Rate (PR), Non Invasive Blood Pressure (NIBP), and ...

Official recall recorded by the U.S. Food & Drug Administration.

The hazard

NIBP valve assembly board subject to this recall includes enhancements that are not included in the current marketing clearance:1. Anti-motion feature 2. NIBP hose blockage identification function

What to do

FDA enforcement status: Terminated

Brands named

mindray ds usa inc dba mindray north americamindraymindray ds

UPCs

11501767900