FDASeptember 27, 2024device

1) Flex Extenders, 5.2" x 15mm OD x 15mm ID REF S6411P 2) Flex Extender, 5.2" x 15mm OD x 22/15mm ID REF 7 5132 The Westmed Flex Extenders are non invasive, single use/disposable, large bore, corrugated plastic tubes that provide a flexible, low resistance lightweight connection between breathing...

Official recall recorded by the U.S. Food & Drug Administration.

The hazard

Due to connector within the circuit flex extender not meeting specifications resulting in a low connection force and as a result, the connector may leak or disconnect and may cause life threating conditions or death.

What to do

FDA enforcement status: Ongoing

Brands named

westmed

UPCs

00709078012488107090780124850070907800840510709078008402