FDAJuly 26, 2019device

Giraffe OmniBed Carestation CS1 - Product Usage: The device can be operated as an incubator or as a warmer and can transition from one mode to the other on user s demand. It cannot be operated in both modes at the same time. Incubators and warmers provide heat in a controlled manner to neonates w...

Official recall recorded by the U.S. Food & Drug Administration.

The hazard

Fasteners potentially torqued to a value less than the specified value and can potentially cause the canopy to become detached and fall.

What to do

FDA enforcement status: Ongoing

Brands named

ge healthcare

Recall history