FDAAugust 8, 2018device

Magellan LeadCare¿ II Blood Lead Test Kit. In Vitro Test for the measurement of lead in fresh whole blood. Catalog Number: 70-6762 Primary DI No: 00850355006000 The LeadCare II Blood Lead Test Kit is for in vitro diagnostic use only. The test kit is for the quantitative measurement of lead in fre...

Official recall recorded by the U.S. Food & Drug Administration.

The hazard

Defect with the calibration button for lot 1808M. Level 1 and Level 2 Quality Control (QC) values will generate results of HIGH (>65 ¿g/dL). Results for patient testing with this lot will be displayed as HIGH.

What to do

FDA enforcement status: Terminated