FDAOctober 9, 2023device

Big Bore RT with software version V4.8.0.10421- Model: 728242 CT scanner and simulator designed for radiation oncology and therapy.

Official recall recorded by the U.S. Food & Drug Administration.

The hazard

For oncology users: If the user performed off-set reconstruction on CT device a shift could be observed on the contouring area of the primary and secondary images sets when using image fusion (on TPS) and incorrect coordinates may be sent to TPS system when using absolute patient marking. This may lead to incorrect radiation therapy planning and possible growth or spread of cancer due to incorrect early treatment regimen. For radiology users: If the user performed off-set reconstruction on CT devices, after multiple acquisitions, the Relate Position could be inaccurate. This could lead to incorrect diagnosis and treatment.

What to do

FDA enforcement status: Ongoing

Brands named

philips north americaphilipsphilips north

UPCs

00884838095168

Recall history