FDAJuly 14, 2017device

Mobile Detector Holder - Stationary X-ray System (RF System). suitable for positioning the wireless portable detector free in the room for X-ray exposures. It supports the wireless portable detector with or without a grid.

Official recall recorded by the U.S. Food & Drug Administration.

The hazard

If the lock mechanism of the movable camp of the mobile detector holder is not adjusted properly or not secured, the portable detector can drop down when the moveable clamp is at the bottom (below of the detector) and additional force is applied. This may happen when the detector is rotated clockwise from landscape to portrait position.

What to do

FDA enforcement status: Terminated

Brands named

philips electronics north americaphilipsphilips electronics