FDAJune 7, 2017device

Brilliance iCT SP (Model No. 728311, Software Version 4.7.0 or 4.7.2 Product Usage: Computed Tomography X-ray systems intended to produce cross-sectional Images of the body by computer reconstruction of X-ray transmission data taken at different angles and planes. These devices may include signal...

Official recall recorded by the U.S. Food & Drug Administration.

The hazard

It was reported that scanners running certain software versions provide confusing information in the "Impressions" section of the Calcium Score report. The generated report provides multiple analyses of the amount of plaque detected, as if they all apply to the patient at the same time. This misleading guidance can be potentially misleading to the user, and may lead to mistreatment of the patient.

What to do

FDA enforcement status: Terminated