VITROS Immunodiagnostic Products HBsAg Controls, Product Code 6800598 For in vitro diagnostic use only. For use in monitoring the performance of the VITROS ECi/ECiQ Immunodiagnostic Systems, the VITROS 3600 Immunodiagnostic System and the VITROS 5600 Integrated System when used for the in vitro q...

Official recall recorded by the U.S. Food & Drug Administration.

The hazard

The signal/cutoff (s/c) results may increase throughout the open-reconstituted timeframe of the allowed 5 days. Increased results may occur on both the negative and positive Controls. The positive Control, although showing elevated results, was still within acceptable performance.

What to do

FDA enforcement status: Terminated