FDASeptember 19, 2016device

Reprocessed Pressure Tourniquet Cuff (PTC) Tourniquet cuffs are single- or dual-bladder inflatable cuffs connected to a tourniquet system via a hose assembly. When wrapped around a limb and inflated, tourniquet cuffs apply an adequate amount of pressure on the arterial blood flow in a limb to cre...

Official recall recorded by the U.S. Food & Drug Administration.

The hazard

Issue with the process for detection of leaking Pressure Tourniquet Cuffs (PTC). Leaking PTC devices may result in excessive blood loss requiring surgical or medical intervention or systemic leakage of local anesthetic during Bier Block requiring medical intervention.

What to do

FDA enforcement status: Terminated

Brands named

stryker sustainability solutionsstrykerstryker sustainability