FDADecember 12, 2017device

DJO EMPOWR KNEE LOCKING TIBIAL PUNCH GUIDE, REF 801-05-042, NON STERILE, Qty 1 For use in orthopedic surgery.

Official recall recorded by the U.S. Food & Drug Administration.

The hazard

The Locking Punch Guide was not sufficiently guiding the large press fit punch, causing the implant to be placed 2-3mm more posterior than intended.

What to do

FDA enforcement status: Terminated

Brands named

encore medical lpencoreencore medical

Recall history

Other federal recalls tied to the same brand — a factual pattern, straight from the source data.

Own something like this?

AllClear screens your whole list against every CPSC, FDA, USDA and NHTSA recall — and tells you the day a new one names something you own.

Start free — 200 items, no card →