FDADecember 12, 2017device

EMPOWR Porous Knee System TIBIAL PUNCH, LARGE, REF 801-05-204, NON STERILE, Qty 1 For use in orthopedic surgery.

Official recall recorded by the U.S. Food & Drug Administration.

The hazard

The Locking Punch Guide was not sufficiently guiding the large press fit punch, causing the implant to be placed 2-3mm more posterior than intended.

What to do

FDA enforcement status: Terminated

Brands named

encore medical lpencoreencore medical